| Voltage Range |
90~264VAC 135~370VDC |
| Input Frequency |
47~63Hz |
| AC Current |
0.8A/100VAC |
| Inrush Current |
35A/230VAC |
| Leakage Current |
0.1mA/240VAC |
| Voltage adj. range |
Fixed |
| Line regulation |
±1.0% |
| Setup, Rise, Hold Up Time |
300mS, 50mS, 16mS at full load |
| Over Load Protection |
0P1J~2P1J: 112~250% rated output power ; 3P1J~8P1J: 150~350% rated output power |
|
Protection type: Hiccup mode, recovers automatically after fault condition is removed |
| Over Voltage Protection |
110~140% rated output voltage |
|
Hiccup mode, recovers automatically after fault condition is removed |
| Over Temperature Protection |
Tj 135°C typically (IC1) detect on main control IC |
|
Will shut down o/p voltage, recovers automatically after temperature goes down |
| Working Temperature |
0°C to +40°C (refer to detating curve) |
| Working Humidity |
20% to 90% RH, non-condensing |
| Storage Temp., humidity |
-20°C to +85°C, 10% to 95% RH |
| Temperature Coefficient |
±0.03%/°C (0°C to 50°C) |
| Vibration |
10 to 500Hz, 2G 10min./1 cycle, period for 60min. each along X, Y, Z axes |
| MTBF |
400Khrs min. MIL-HDBK-217F (25°C) |
| Safety Standards |
UL2601-1, IEC60601-1, EN60601-1 Approved |
| Withstand Voltage |
I/P-O/P: 5656VDC |
| Isolation Resistance |
I/P-O/P: 100M Ohms/ 500VDC |
| EMI Conduction & Radiation |
Compliance to EN55011 (CISPR11) Class B |
| Harmonic Current |
Compliance to EN61000-3-2,3 |
| EMS Immunity |
Compliance to EN60601-1-2 (EN61000-4-2,3,4,5,6,8,11) Light industry level, criteria A |
| Connector plug |
Standard type P1J:2.1mm*5.5mm*11mm, turning fork type, center positive |
| Connector cable |
Standard type UL1185 6ft (with ferrite core) |