MPS-65-3-3 - MPS-65-3.3
Home > Products > MPS-65 Series 65W Single Output Medical Open Frame Power Supply > MPS-65-3.3
|
Standards
|
General Specifications
| MAXIMUM OUTPUT POWER |
72W (+3.3V:50W;+5V:69W) WITH 18 CFM MIN. FORCED AIR CONVECTION |
| DC VOLTAGE ADJ. |
?10% |
| INPUT VOLTAGE RANGE |
90~264VAC 47~440Hz ; 127~370VDC |
| AC CURRENT |
1.6A/115V 0.9A/230V |
| INRUSH CURRENT |
COLD START 20A/115V 40A/230V |
| LEAKAGE CURRENT |
<0.3mA/264VAC |
| OVERLOAD PROTECTION |
OVERLOAD PROTECTION73~105W (3.3V:51~75W)(5V:70~105W) TYPE:PULSING HICCUP SHUTDOWN RESET:AUTORECO |
| OVER VOLTAGE PROTECTION |
115%~135% |
| TEMP. COEFFICIENT |
±0.04% / °C (0~50°C) |
| SETUP, RISE, HOLD UP TIME |
800ms, 30ms, 20ms/230VAC 800ms, 30ms, 12ms/115VAC |
| VIBRATION |
10~500Hz, 2G 10min./1cycle, PERIOD FOR 60min. EACH AXES |
| WITHSTAND VOLTAGE |
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC |
| ISOLATION RESISTANCE |
I/P-O/P, I/P-FG, O/P-FG:500VDC / 100M Ohms |
| WORKING TEMP., HUMIDITY |
-10°C~+60°C(REFER TO OUTPUT DERATING CURVE), 20%~90% RH |
| STORAGE TEMP., HUMIDITY |
-20°C~+85°C, 10%~95% RH |
| SAFETY STANDARDS |
UL2601-1, EN60601-1 APPROVED |
| EMC |
CISPR11 (EN55011), IEC61000-4-2,3,4,5,6,8,11 IEC61000-3-2,3EN60601-1-2 VERIFICATION |
|
MPS-65-3.3
|
DC Voltage
|
Rated Current
|
Current Range
|
Rated Power
|
Ripple & Noise
|
Voltage ADJ Range
|
Voltage Tolerance
|
Line Regulation
|
Load Regulation
|
Efficiency
|
Over Voltage
|
|
3.3V
|
12A
|
0-15.2A
|
39.6W
|
80mVp-p
|
2.97-3.63V
|
±3.0%
|
±1.0%
|
±3.0%
|
66%
|
3.8-4.46V
|
|
 |
| Add this product to your ClipBook |