MPT-45A - MPT-45A
Home > Products > Medical Power Supply > MPT-45 Series 45W Triple Output Medical Open Frame Power Supply > MPT-45A
General Specifications
| MAXIMUM OUTPUT POWER |
52W WITH 18 CFM MIN. FORCED AIR CONVECTION |
| INPUT VOLTAGE RANGE |
90~264VAC 47~440Hz ; 127~370VDC |
| AC CURRENT |
1.2A/115V 0.7A/230V |
| INRUSH CURRENT |
COLD START 20A/115V 40A/230V |
| LEAKAGE CURRENT |
<0.3mA/264VAC |
| OVERLOAD PROTECTION |
53~75W OUTPUT POWER TYPE:PULSING HICCUP SHUTDOWN RESET:AUTO RECOVERY |
| OVER VOLTAGE PROTECTION |
5.75~6.75VDC ON CH1 |
| TEMP. COEFFICIENT |
±0.04% / °C (0~50°C) ON +5V OUTPUT |
| SETUP,RISE,HOLD UP TIME |
800ms, 20ms, 20ms/230VAC 800ms, 20ms, 12ms/115VAC |
| VIBRATION |
10~500Hz, 2G 3AXES 10min. / 1cycle ( 1 HOUR / EACH AXES ) |
| WITHSTAND VOLTAGE |
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC 1min. |
| ISOLATION RESISTANCE |
I/P-O/P,I/P-FG,O/P-FG:500VDC / 100M Ohms |
| WORKING TEMP., HUMIDITY |
-10°C~+60°C(REFER TO OUTPUT DERATING CURVE), 20%~90% RH |
| STORAGE TEMP., HUMIDITY |
-20°C~+85°C, 10%~95% RH |
| SAFETY STANDARDS |
UL2601-1, EN60601-1 APPROVED |
| EMC |
CISPR11 (EN55011), IEC61000-4-2,3,4,5,6,8,11 IEC61000-3-2,3EN60601-1-2 VERIFICATION |
MPT-45A Specifications
| Part Number |
Voltage (max/rated) | Current (amps) | Power (watts) | R&N (mV) | Efficiency (%) |
| MPT-45A
|
5V
|
3A
|
40.05W
|
60mV
|
73%
|
|
|
12V
|
2A
|
|
120mV
|
|
|
|
-5V
|
0.3A
|
|
60mV
|
|
MPT-45A
| DC Voltage | Rated Current | Current Range | Rated Power | Ripple & Noise | Voltage Tolerance | Line Regulation | Load Regulation | Efficiency |
| 5V | 3A | 0.4-5A | 40.5W | 60mVp-p | ±4.0% | ±1.0% | ±3.0% | 73% |
| 12V | 2A | 0.2-2.5A | 40.5W | 120mVp-p | ±7.0% | ±2.0% | ±4.0% | 73% |
| -5V | 0.3A | 0-0.5A | 40.5W | 60mVp-p | ±5.0% | ±1.0% | ±1.0% | 73% |