Specification
| AC input voltage |
90~264 VAC, 47~440Hz / 127~370VDC |
| AC input current |
Maximum input current 2.9A at 115VAC, 60Hz or 1.7A at 230VAC, 60Hz with 100% output load. |
| Inrush current |
Less than 22A at 115VAC or 45A at 230VAC at cold start conditions |
| Setup, rise time |
500ms, 20ms / 230VAC at full load, 1200ms, 50ms / 115VAC at full load |
| Hold-up tiime |
80ms / 230VAC at full load,14ms / 115VAC at full load |
| Leakage current |
Leakage current is less than 180µA at 264VAC |
| DC output adj. range |
Between -5%~+10% rated output voltage |
| Over load protection |
Will be activated at 110~150% rated load. |
| Over voltage protection |
115% ~ 135% rated output voltage |
| Power good/Fail signal (optional) |
At 90% of rated value, a+5V TTL signal will be sent out with a 10~500ms delay; at least |
|
1ms before the output voltage goes below 90% of the rated value,the TTL signal will be turned off. |
| Working temp. |
operate from -20~70°C . |
| Workinig humidity |
20~90% RH non-condensing. |
| Storage temp., humidity |
-40~+85 °C, 10~90% RH |
| Temp. coefficient |
±0.04%/°C on all outputs at full load between 0~50°C of ambient temperature. |
| Vibration |
2G of acceleration,vibrating frequency adjust from 10Hz ~500Hz within a 10 min cycles |
| Safety standards |
Medical:UL60601-1, TUV EN60601-1 approved |
|
Commercial: Also design refer to UL60950-1, TUV EN60950-1 |
| Withstand voltage |
4000VAC between input and output,1500VAC between input and F.G.,500VAC between output and F.G. |
| Isolation resistance |
>100M Ohms for I/P-O/P, I/P-FG, O/P-FG by using 500VDC test voltage. |
| EMI compliance |
Compliance Level EN55011, Class B,EN55022, Class B,EN61000-3-2,EN61000-3-3 |
| EMS compliance |
Compliance Level EN61000-4-2,Level 3, 8KV,EN61000-4-2 Level 2, 4KV |
|
EN61000-4-3 Level 2, 3V/m,Level 3, 10V/m,EN61000-4-4 Level 2, 1KV/5KHz,Level 3, 2KV/5KHz |
|
EN61000-4-5,Level 4, 2KV/Line-Line 4KV/Line-Earth,EN61000-4-6 Level 2, 3Vrms/m,Level 3, 10Vrms/m |
|
EN61000-4-8,Level 2, 3A/m,Level 3, 10A/m,EN61000-4-11 Compliance |
|
ENV50204,Level 2, 3V/m, 900MHz,Level 3, 10A/m, 900MHz |
| MTBF |
262,100 hours min at full load and 25°C of ambient temperature. |
|