MEAN WELL Products
HomeMEAN WELL ProductsOpen Frame Switching Power Supply160W Dual Output & 5Vsb Open Frame Power Supply
Description
RPDG-160 Series, triple output medical power supply with additional PS-ON control, 5V/0.8A standby output and 0.75W of no-load-power-consumption can be implemented into all kinds of end systems.
Additional information
Standards | CB, CE, CSUUS, PFC, MEDICAL, TUVMedical |
---|---|
Type | AC/DC |
Outputs | Dual |
- Universal AC input
- Built-in active PFC function
- Short circuit, over load, over voltage, over temperature protections
- 5"x3" compact size
- Free air convection for 100W and 150W with 20.5 CFM forced air
- With power good and fail signal output
- No load power consumption under 0.75W by PS-ON control
- Standby 5V@0.8A with Fan/0.6A without Fan
- Approval: UL / CUL / TUV / CE / CB
- Medical applications
- IT applications
Input Voltage | 90 ~ 264Vac 127 ~ 370Vdc |
Input Frequency | 47 ~ 63Hz |
Power Factor | PF>0.93/230Vac PF>0.98/115Vac at full load |
Efficiency | 84% |
Input Current | 1.7A/115Vac 0.9A/230Vac |
Inrush Current | 35A/115Vac 70A/230Vac, cold start |
Leakage Current | Earth leakage current <300uA/ 264Vac, Patient leakage current <100uA/264Vac |
Rated Power (Convection) | 100.2W |
Rated Power (20.5CFM) | 150.4W |
Set Up, Rise Time | 1200ms, 30ms/230Vac 2500ms, 30ms/115Vac at full load |
Hold Up Time | 16ms/230Vac/115Vac at full load |
Over Load Protection | 105 ~ 135% rated output power |
Hiccup mode, recovers automatically after fault condition is removed | |
Over Voltage Protection | CH1 Only |
Shut down o/p voltage, re-power on to recover | |
Over Temperature Protection | 105°C (TSW1)detect on heatsink of power transistor |
90°C (TSW2 )detect on heatsink of power transistor | |
(TSW1) Shut down o/p voltage, recovers automatically after temperature goes down | |
(TSW2) Shut down o/p voltage, re-power on to recover | |
5V Standby | 5VSB : 5V@0.6A without Fan, 0.8A with Fan 20.5CFM; tolerance ±2%, ripple : 50mVp-p (max.) |
PS-ON Input Signal | Power on: PS-ON = Hi or >2V ; Power off: PS-ON = Low or <0.5V |
Power Good/Power Fail | 500ms>PG>10ms PF>1ms |
Working Temperature | -20°C ~ +70°C (Refer to output load derating curve) |
Working Humidity | 20 ~ 90% RH non-condensing |
Storage Temperature, Humidity | -40°C ~ +85°C, 10 ~ 95% RH |
Temperature Coefficient | ±0.03%/°C (0 ~ 50°C) |
Vibration | 10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes |
Withstand Voltage | I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC |
Isolation Resistance | I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC/ 25/°C 70% RH |
Safety Standards | UL60601-1, TUV EN60601-1 approved |
EMC | Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B; EN61000-3-2,-3; ENV50204; |
EN61000-4-2,3,4,5,6,8,11; EN55024, EN60601-1-2, EN61204-3, medical level, criteria A | |
MTBF | 196.3Khrs min. MIL-HDBK-217F (25°C) |
Dimensions | 127 x 76.2 x 34.6mm |
- CB
- CE
- CSUUS
- MEDICAL
- PFC
- TUVMedical
Products in this range
Part number | Power (watts) | Voltage (max/rated) | Current (amps) | R&N (mV) | Efficiency (%) | |
---|---|---|---|---|---|---|
RPDG-160B | 160W | 24V | 4A | 200mV | 84% | Request Quote |