?> ?> ?> ?> ?> MPT-120 Series - 120W Triple Output Open Frame Medical Power Supply
escort bayan fethiye kuşadası escort bayan escort bayan bodrum

MEAN WELL Products

HomeMEAN WELL ProductsMedical Power Supply120W Triple Output Open Frame Medical Power Supply

MPT-120 Series - 120W Triple Output Open Frame Medical Power Supply

Email

Additional information

Standards

CB, CE, CSUUS, MEDICAL, TUVMedical

Type

AC/DC

Outputs

Triple

  • Universal AC input / Full range
  • Low leakage current <180µA
  • Protections:Short circuit/Over load/Over voltage
  • UL60601-1 medical safety approved
  • With power good and fail signal output (Optional)
  • 100% full load burn-in test
  • Fixed switching frequency at 45KHz
AC input voltage90~264 VAC, 47~440Hz / 127~370VDC
AC input currentMaximum input current 2.9A at 115VAC, 60Hz or 1.7A at 230VAC, 60Hz with 100% output load.
Inrush currentLess than 22A at 115VAC or 45A at 230VAC at cold start conditions
Setup, rise time500ms, 20ms / 230VAC at full load, 1200ms, 50ms / 115VAC at full load
Hold-up tiime80ms / 230VAC at full load,14ms / 115VAC at full load
Leakage currentLeakage current is less than 180µA at 264VAC
DC output adj. rangeBetween -5%~+10% rated output voltage
Over load protectionWill be activated at 110~150% rated load.
Over voltage protection115% ~ 135% rated output voltage
Power good/Fail signal (optional)At 90% of rated value, a+5V TTL signal will be sent out with a 10~500ms delay; at least
1ms before the output voltage goes below 90% of the rated value,the TTL signal will be turned off.
Working temp.operate from -20~70°C .
Workinig humidity20~90% RH non-condensing.
Storage temp., humidity-40~+85 °C, 10~90% RH
Temp. coefficient±0.04%/°C on all outputs at full load between 0~50°C of ambient temperature.
Vibration2G of acceleration,vibrating frequency adjust from 10Hz ~500Hz within a 10 min cycles
Safety standardsMedical:UL60601-1, TUV EN60601-1 approved
Commercial: Also design refer to UL60950-1, TUV EN60950-1
Withstand voltage4000VAC between input and output,1500VAC between input and F.G.,500VAC between output and F.G.
Isolation resistance>100M Ohms for I/P-O/P, I/P-FG, O/P-FG by using 500VDC test voltage.
EMI complianceCompliance Level EN55011, Class B,EN55022, Class B,EN61000-3-2,EN61000-3-3
EMS complianceCompliance Level EN61000-4-2,Level 3, 8KV,EN61000-4-2 Level 2, 4KV
EN61000-4-3 Level 2, 3V/m,Level 3, 10V/m,EN61000-4-4 Level 2, 1KV/5KHz,Level 3, 2KV/5KHz
EN61000-4-5,Level 4, 2KV/Line-Line 4KV/Line-Earth,EN61000-4-6 Level 2, 3Vrms/m,Level 3, 10Vrms/m
EN61000-4-8,Level 2, 3A/m,Level 3, 10A/m,EN61000-4-11 Compliance
ENV50204,Level 2, 3V/m, 900MHz,Level 3, 10A/m, 900MHz
MTBF262,100 hours min at full load and 25°C of ambient temperature.
  • CB
  • CE
  • CSUUS
  • MEDICAL
  • TUVMedical